WDH-1 (A)
Mode: A/C, A/C+SIGH,SIMV,SIMV+PSV,PCV,PSV, MANUAL , PEEP,CPAP
Display: 10.4" TFT color screen
Operation method: knob operation and touch key
Ventilation wave type: squire wave, deceleration wave
Tidal volume: 50 - 1500ml
Breath frequency: 2 - 60 BPM
I :E ratio: 3:1 - 1:3
Inspiration time: 0.2 - 3 S
Inspiration plateau: 0 - 50%
PEEP: 0 - 2.0 Kpa
Trigger sensitivity: -0.1 - -2.0 KPa, OFF
Oxygen concentration: 45- 100%
PSV pressure: 0 - 3.5 KPa
Monitor wave and parameters:
Monitor wave: P-t, V-t Tidal volume
Breath frequency I:E ratio PEEP Ppeak
MV Oxygen concentration Pmean
Alarm parameters and alarm setting:
Tidal volume upper limit: 100 - 2000 ml
Tidal volume lower limit: 20 - 2000 ml
Breath frequency upper limit: 5 - 100 bpm
Breath frequency lower limit: 2 - 60 BPM
Peak pressure upper limit: 0.5 - 7.0 KPa
Peak pressure lower limit: 0 - 3.0 Kpa
PEEP upper limit: 0.5 - 3.0 KPa
PEEP lower limit: 0 - 2.0 KPa
Oxygen concentration upper limit: 30% - 100%
Oxygen concentration lower limit: 21% - 60%
MV upper limit: 4.0 - 40 L/min
MV lower limit: 0.5 - 20 L/min
Apnea alarm and setting: 3 - 12 S
Battery low alarm: DC 11.5V
Other function:
Build-in battery